Insulation
► Nominal test voltages: 250 V, 500 V, 1000 V
► Insulation resistance range: 20 GΩ
Range Full Scale Accuracy
► 1000 volts ±3% ±2 digits ±0.2% per GΩ
► 500 volts. ±3% ±2 digits ±0.4% per GΩ
► 250 volts. ±3% ±2 digits ±0.8% per GΩ
► Analogue range: 1 GΩ full scale
► Short Circuit Current: 2 mA +0% -50%
► Terminal voltage: -0% +20% ±1 V
► Test Current on load: 1 mA at min. pass value of insulation specified in BS7671, HD384 and IEC364, 2 mA max.
► EN61557 Operating range: 0,10 MΩ to 1,00 GΩ
► Leakage Current: 10% ±3 digits
► Voltage display: 3% ±3 digits ±0.5% of rated voltage
► Polarisation Index (PI): 10 min / 1minute ratio
► Dielectric Absorption Ratio (DAR): 60 sec / 30 sec ratio
Notes:
(1) All ranges measure from 0,00 MΩ upwards.
(2) Above specifications only apply when high quality silicone leads are being used.
Continuity
► Measurement: 0,01 Ω to 99,9 Ω (0 to 100 Ω on analogue scale)
► Accuracy: ±3% ±2 digits (0 to 100 Ω)
► Open circuit voltage: 5 V ±1 V
Test current:
► 200 mA (–0 mA +20 mA) (0.01 Ω to 9.99 Ω)
► 20mA (±1 mA ) (10.0 Ω to 99.9 Ω)
► Zero offset at probe tips: 0,10 Ω typical
► Lead resistance zeroing: Up to 9,00 Ω
► Buzzer: Variable limit 1 Ω, 2 Ω, 5 Ω, 10 Ω, 20 Ω
Resistance
► Measurement: 0,01 kΩ to 1000 kΩ (0 to 1 MΩ on analogue scale)
► Accuracy: ±3% ±2 digits
► Open circuit voltage: 5 V ±1 V
► Short circuit current: 20 μA ± 5 μA
Voltage range
► 0 to 600 V d.c. ±2% ±2 digits
► 10 mV to 600 V TRMS sinusoidal (40 to 400 Hz) ±2% ±2 digits
► 0 to 1000 V on Analogue scale
► Unspecified input level 0 - 10 mV (40 to 400 Hz)
For non-sinusoidal waveforms additional specification apply:
►±3% ±2 digits 101 mV to 600 V TRMS and ±8% ±2 digits 10 mV to 100 mV TRMS
► Default Voltmeter: Operates at >25 V a.c. or d.c. on any range except OFF
Frequency:
► 40-450 Hz (40 Hz - 99,9 Hz)
► ±0.5% ±1 digit (100 Hz to 450 Hz)
► Power Supply: 5 x 1,5 V cells type IEC LR6 (AA, MN1500, HP7, AM3 R6HP) Alkaline NiMH rechargeable cells may be used.
► Battery life: 2200 insulation tests with duty cycle of 5 sec ON /55 sec OFF @ 1000 V into 1 MΩ
Dimensions
► Instrument: 220 x 92 x 50mm
► Instrument + case: 456 x 178 x 89mm
Weight
► Instrument only: 590 g, 775 g with boot
► Instrument plus case: 1.75kg
Fuse
► Use only a 500 mA (FF) 1000 V 32 x 6 mm ceramic fuse of high breaking capacity HBC 50 kA minimum. Glass fuses MUST NOT be fitted.
Safety Protection
► The instruments meet EN 61010-1 (1995) to 600 V phase to earth, Category IV. Refer to safety warnings supplied
E.M.C.
► In accordance with IEC 61326 including amendment No.1
Environmental
► Operating range: -10 to +55°C
► Operating humidity: 90% RH at 40°C max.
► Storage temperature range: -25 to +70°C
► Maximum altitude: 2000 m
► Dust and water protection: IP54 Protected against dust and splashing water