Products in: High Voltage Test Equipment/AC Test Systems/  T&R KV30-40 mk2 High Voltage AC Test System

 T&R KV30-40 mk2 High Voltage AC Test System

T&R KV30-40 mk2 High Voltage AC Test System
Now £5549.00
Was £5967.00
Save £418.00
Buy Now
Variable High Voltage output 0 to 30kV 40mA

Be the first to review this product!

Product Features: T&R KV30-40 mk2 High Voltage AC Test System

T&R KV30-40 Mk2 AC Test Set

  •  
  • 0-30kV AC (KV30-40 model) output voltage
  • 1.2kVA output capability
  • Key operated supply switch to prevent unauthorised operation
  • Dual overload protection
  • Variable electronic trip - 10-110% of rated output
  • Voltage and current metering
  • Zero-volt interlock
  • External 24Vdc interlock circuit
  • Visual indication of test piece failure

    The T and R KV30-40 Mk2 series high voltage AC test sets are general purpose electrical test instruments designed for testing insulation systems and the measurement of breakdown voltage on electrical plant and components.

    These High voltage T and R test sets have an AC output voltage variable from zero to 10kV (T and R KV10-120 model), 15kV (T and R KV15-80 model) or 30kV (T and R KV30-40 model) model with accurate metering on both the output voltage and current. For safety a zero-voltage interlock is provided, ensuring that the output may only be energised with the voltage control at zero.

    The AC output voltage is metered by a large, linear analogue instrument marked 0-12kV, 0-20kV and 0-40kV on the three units. All units have a x0.5 range on the voltmeter giving a full scale of 6kV ,10kV or 20kV respectively. Load current is metered by a second analogue instrument with a full scale of 140mA (KV10-120), 100mA (KV15-80) or 50mA (KV30-40). A 10 range is provided on both units, allowing the full scale on the current metering to be reduced by a factor of 10.

    The test object and output are protected by an adjustable current trip. The trip level may be set to 10-110% of the rated output in 10% steps. Breakdown of the test object is visibly indicated. The unit must be manually reset after a trip condition before testing can be resumed.

    The units feature automatic mains voltage selection and an external 24Vdc interlock circuit for connection of emergency off switches aned enclosure interlocks.

Supplied Accessories: T&R KV30-40 mk2 High Voltage AC Test System

The unit is supplied in a compact, portable case with a permanently connected 2 metre long HV output lead. A T&R ES30 earthing stick is also supplied as standard. You also get spare fuses, earth lead and Operatinig Manual.
 

Product Specifications: T&R KV30-40 mk2 High Voltage AC Test System

Output
The output of the unit is by a permanently connected partially screened high voltage cable.
Maximum voltage 30kV AC, Maximum current 40mA

Unit Type Voltage Continuous 5 minutes
KV30-40 0 to 30kV 20mA 40mA  
 

The above intermittent on times must be followed by an off time of 15 minutes, and are based on an ambient temperature of 25°C.

 

Metering
The output voltage on the KV30-40 is metered by a dual range analogue instrument. 20kV / 40kV

Unit

Range

Accuracy  
KV30-40 0 to 40kV ±2% of full scale  
 
The output current on the KV30-40 is metered by a dual range analogue instrument. 5mA / 50 mA full scale
 

Supply Requirements
115V/230V±10% auto-selecting
50/60Hz 1ph 1200VA max


Overload Protection

Two overload protection circuits are provided on the units. The first is user selectable, and allows trip currents between 10% and
110% of the rated output to be set. A trip condition is indicated by an illuminated push button and an audible alarm. The second
trip circuit is a magnetic circuit breaker operating on the primary of the HV transformer. This operates on large overloads (such
as under flashovers).

Protection and Safety
In addition to the output protection the input and control supplies are protected by fuses.
All units are designed to meet the requirements of BS EN61010.
An earth terminal is provided on the transformer which must be connected to a low impedance local earth.

Recommended Accessories: T&R KV30-40 mk2 High Voltage AC Test System

Submit Review

Please complete the information below and it will be displayed on our website once it has been reviewed by a member of staff.

Contribution type:

Rating:

Spambot:

Name:

Email: (Not displayed)

Review Title:

Review / Comment Details:



test-meter.co.uk - Home | Your Order | Latest Special Offers | Terms & Conditions | Contact Us | Request Link | Links | Site Map

Prices exclude delivery carriage at £10 and VAT at 20%

All brand names are trade marked to their respective owners.

Company Number: 5394680 VAT Number: GB 842 7016 40


test-meter.co.uk
is a trading name of PAT Training Services Limited.
PAT Training Services Limited.
1 Aston Court,
Town End Close,
LEEDS,
LS13 2AF
UK

Tel: 0113 248 9966
Fax: 0113 248 4276

admin@test-meter.co.uk

Copyright © PAT Training Services Limited, All rights reserved. E&OE