Output
The T & R PT30-10s cable test systems have high quality high voltage output connectors, and are supplied with detachable, partially screened output cables
Test Set Voltage Continuous 5 minutes
PT30-10 0 to ±30kV 5mA 10mA
The above intermittent on times must be followed by an off time of 15 minutes, and is based on an ambient temperature of 25°C.
Metering
The positive and negative output voltages on the PT series are metered on the HV output by separate analogue instruments.
Test Set Voltage Range Accuracy
PT30-10 0-30kV ±1.5% of full scale
The output current on both of the outputs is metered by a dual range analogue instrument. The 1mA range is selected by the¸10 pushbutton adjacent to the mA meter.
Test Set Current Range Accuracy
PT30-10 10mA ±2.5% of full scale
1mA ±2.5% of full scale
Overload Protection
Protected by an overload trip on the output that operates at 12mA.
Load Discharge System
The PT30-10 is fitted with an automatic internal load discharge system that grounds the load via a 20kW resistor on each output when the output is switched off. The discharge system is rated to dissipate 2.5kJ once every 15 minutes on each output. The PT30-10 can discharge a maximum load capacitance of 4mF per output from 30kV.
Maximum discharge energy 2.5kJ 2.5kJ
Maximum discharge capacitance 10mF 4mF from unit max output voltage
Supplied with a DP40 manual discharge probe to allow the discharge of higher capacitance loads. The DP40 discharge probes consist of a pointed probe connected to a 5m long earth lead via a series of surge resistors with an insulated handle. The earth lead insulation is clear silicone allowing the conductor to be easily inspected, and is terminated in an M6 hooked crimp.
Discharge probe max discharge voltage 40kV
Resistance 60kW
Maximum discharge energy 7.2kJ
Maximum discharge capacitance 11mF
from DP max rated voltage
Supply Requirements
PT30-10 mk2 115V/230V±10% auto-selecting 50/60Hz 1ph 750VA max
Protection and Safety
The output of the unit is protected by an overload trip, and the input and control supplies are protected by fuses.
The PT30-10 is designed to meet the requirements of BS EN61010.
An earth terminal is provided on the units which must be connected to a low impedance local earth.
Temperature Range
Storage -20°C to 60°C Operating 0°C to 45°C
Dimensions and Weight
T and R PT30-10 471x191x362mm 25kg