Specification
Continuity // Protective earth resistance Continuity - Ranges: 0.00 Ω … 19.99 Ω / 20.0 Ω … 99.9 Ω / 100.0 Ω … 199.9 Ω / 200 Ω … 999 Ω
- Resolution: 0.01 Ω / 0.1 Ω / 0.1 Ω / 1 Ω
- Accuracy: ±(2 % of reading + 2 D) / ± 3 % of reading / ± 5 % of reading / indicative
- Operating range (acc. to EN 61557-4)............. 0.08 Ω … 199.9 Ω
- Test currents..................................................... 0.2 A,10 A, 25 A
- Current source (at nominal mains voltage, use of standard accessories)
- .......................................................................... > 0.2 A at R < 2 Ω
- .......................................................................... > 10 A at R < 0.1 Ω at 230 V
- .......................................................................... > 25 A into short circuit at 230 V
- Open circuit voltage.......................................... < 9 V a.c.
Test terminals: Continuity - R (200 mA): P/S – PE, Socket PE; Socket PE – IEC PE
- R (10 A, 25 A): P/S – Socket PE; Socket PE – IEC PE
Protective earth resistance - R (200 mA): P/AP – PE, Socket PE
- R (10 A, 25 A): P/AP – Socket PE
Insulation Resistance (Riso, Riso-S) - Insulation resistance, Insulation resistance –S (250 V, 500 V)
- Range: 0.00 MΩ … 19.99 MΩ / 20.0 MΩ … 99.9 MΩ / 100.0 MΩ … 199.9 MΩ
- Resolution: 0.01 MΩ / 0.1 MΩ / 0.1 MΩ
- Accuracy: ±(3 % of reading + 2 D) / ± 5 % of reading /± 10 % of reading
Output voltage (Um) - Range: 0 V … 600 V
- Resolution: 1V
- Accuracy: ±(3 % of reading + 2 D)
- Operating range (acc. to EN 61557-2).........0.08 MΩ … 199.9 (999) MΩ
- Nominal voltages Un ....................................250 V, 500 V (- 0 %, + 10 %)
- Short circuit current ......................................max. 2.0 mA
Test terminals: - Riso LN, Socket LN – PE, Socket PE, P/S
- Riso-S LN, Socket LN – P/S
Sub-Leakage Current, Substitute Leakage Current - S (Isub, Isub-S) - Range: 0.00 mA … 1.99 mA / 2.00 mA … 19.99 mA
- Resolution: 0.01mA / 0.01mA
- Accuracy: ±(3 % of reading + 3 D) / ±(5 % of reading)
- Operating range (acc. to EN 61557-16)........... 0.02 mA … 19.99 mA
- Open circuit voltage.......................................... 230 V a.c., 110 V a.c.
- Current calculated to mains supply voltage (110 V or 230 V) is displayed.
Test terminals: - Isub LN, Socket LN – PE, Socket PE, P/S
- Isub-S LN, Socket LN – P/S
Differential leakage current (Idiff) Differential leakage current - Range: 0.000 mA … 1.999 mA / 2.00 mA … 19.99 mA
- Resolution: 1 µA / 0.01 mA
- Accuracy: ±(3 % of reading + 3 D) / ±(5 % of reading)
Power (active) (P) - Range: 0 W … 999 W / 1.00 kW … 3.70 kW
- Resolution: 1W / 10W
- Accuracy: ±(5 % of reading + 5 D) / ±5 % of reading
- Operating range (acc. to EN 61557-16)........... 0.010 mA … 19.99 mA
- Influence of load current................................... < 0.02 mA/A
Test terminals: - Idiff Socket L,N – Socket PE, P/S
- P Socket L – Socket N
PE leakage current (Ipe) PE leakage current - Range: 0.000 mA … 1.999 mA / 2.00 mA … 19.99 mA
- Resolution: 1µA / 0.01mA
- Accuracy: ±(3 % of reading + 3 D) / ±(5 % of reading)
Power (active) (P) - Range: 0 W … 999 W / 1.00 kW … 3.70 kW
- Resolution: 1W / 10W
- Accuracy: ±(5 % of reading + 5 D) / ±5 % of reading
- Operating range (acc. to EN 61557-16)........... 0.010 mA … 19.99 mA
Test terminals: - Ipe Socket L,N – Socket PE
- P Socket L – Socket N
Touch leakage current (Ipe) Touch leakage current - Range: 0.000 mA … 1.999 mA / 2.00 mA … 19.99 mA
- Resolution: 1 µA / 0.01mA
- Accuracy: ±(3 % of reading + 3 D) / ±(5 % of reading)
Power (active) (P) - Range: 0 W … 999 W / 1.00 kW … 3.70 kW
- Resolution: 1W / 10W
- Accuracy: ±(5 % of reading + 5 D) / ±5 % of reading
- Operating range (acc. to EN 61557-16)........... 0.010 mA … 19.99 mA
Test terminals: - Itou: Socket L,N – P/S
- P: Socket L – Socket N
Power Power (active) (P) - Range: 0 W … 999 W / 1.00 kW … 3.70 kW
- Resolution: 1W / 10W
- Accuracy: ±(5 % of reading + 5 D) / ±5 % of reading
Power (apparent) (S) - Range: 0 VA … 999 VA / 1.00 kVA … 3.70 kVA
- Resolution: 1 VA / 10VA
- Accuracy: ±(5 % of reading + 5 D) / ±5 % of reading
Power (reactive) (Q) - Range: ±(0 VAr … 999) VAr / ±(1.00 kVAr … 3.70) kVAr
- Resolution: 1 VAr / 10 VAr
- Accuracy: ±(5 % of reading + 5 D) / ±5 % of reading
Power factor (PF) - Range 0.00i … 1.00i / 0.00c … 1.00c
- Resolution: 0.01
- Accuracy: ±(5 % of reading + 5 D)
Total Harmonic Distortion (voltage) (THDU) - Range: 0.0 % … 99.9 %
- Resolution: 0.1 %
- Accuracy: ±(5 % of reading + 5 D)
Total Harmonic Distortion (current) (THDI) - Range: 0.00 A … 16.00 A
- Resolution: 0.01 A
- Accuracy: ±(3 % of reading + 5 D)
Cosinus Φ (Cos Φ) - Range: 0.00i … 1.00i / 0.00c … 1.00c
- Resolution: 0.01
- Accuracy: ±(5 % of reading + 5 D)
Current (I) - Range: 0.00 A … 16.00 A
- Resolution: 0.01 A
- Accuracy: ±(3 % of reading + 5 D)
Voltage (U) - Range: 0.0 V … 199.9 V / 200 V … 264 V
- Resolution: 0.1V / 1V
- Accuracy: ±(3 % of reading + 10 D) / ±3 % of reading
Test terminals: - P,S,Q,PF,THDU,THDI, Cos Φ, I, U: Socket L – Socket N
Leaks & Power Power (active) (P) - Range: 0 W … 999 W / 1.00 kW … 3.70 kW
- Resolution: 1W / 10W
- Accuracy: ±(5 % of reading + 5 D) / ±5 % of reading
Touch leakage current: (Itou) - Range: 0.000 mA … 1.999 mA / 2.00 mA … 19.99 mA
- Resolution: 1 µA / 0.01 mA
- Accuracy: ±(3 % of reading + 3 D) / ±(5 % of reading)
- Operating range (acc. to EN 61557-16)........... 0.010 mA … 19.99 mA
Differential leakage current (Idiff) - Range: 0.000 mA … 1.999 mA / 2.00 mA … 19.99 mA
- Resolution: 1 µA / 0.01 mA
- Accuracy: ±(3 % of reading + 3 D) / ±(5 % of reading)
- Operating range (acc. to EN 61557-16)........... 0.010 mA … 19.99 mA
- Influence of load current................................... < 0.02 mA/A
Power (apparent) (S) - Range: 0 VA … 999 VA / 1.00 kVA … 3.70 kVA
- Resolution: 1 VA / 10VA
- Accuracy: ±(5 % of reading + 5 D) / ±5 % of reading
Power (reactive) (Q) - Range: ±(0 VAr … 999) VAr / ±(1.00 kVAr … 3.70) kVAr
- Resolution: 1 VAr / 10 VAr
- Accuracy: ±(5 % of reading + 5 D) / ±5 % of reading
Power factor (PF) - Range 0.00i … 1.00i / 0.00c … 1.00c
- Resolution: 0.01
- Accuracy: ±(5 % of reading + 5 D)
Total Harmonic Distortion (voltage) (THDU) - Range: 0.0 % … 99.9 %
- Resolution: 0.1 %
- Accuracy: ±(5 % of reading + 5 D)
Total Harmonic Distortion (current) (THDI) - Range: 0.00 A … 16.00 A
- Resolution: 0.01 A
- Accuracy: ±(3 % of reading + 5 D)
Cosinus Φ (Cos Φ) - Range: 0.00i … 1.00i / 0.00c … 1.00c
- Resolution: 0.01
- Accuracy: ±(5 % of reading + 5 D)
Current (I) - Range: 0.00 A … 16.00 A
- Resolution: 0.01 A
- Accuracy: ±(3 % of reading + 5 D)
Voltage (U) - Range: 0.0 V … 199.9 V / 200 V … 264 V
- Resolution: 0.1V / 1V
- Accuracy: ±(3 % of reading + 10 D) / ±3 % of reading
Test terminals: - P,S,Q,PF,THDU,THDI, Cos Φ, I, U: Socket L – Socket N
- Idiff: Socket L,N – Socket PE, P/S
- Itou: Socket L,N – P/S
PRCD test Trip-out time (tΔN) - Range: 0 ms … 300 ms (999 ms*) (½xI∆N) / 0 ms … 300 ms (I∆N) / 0 ms … 40 ms (5xI∆N)
- Resolution: 1ms / 1ms / 1ms
- Accuracy: ±3ms / ±3ms / ±3ms
*According to standard AS/NZS 3017 - Trip out-current (IΔ)
- Range: 0.2xI∆N … 2.2xI∆N
- Resolution: 0.05xI∆N
- Accuracy: ±0.1xI∆N
- Test current type:.................sine-wave (AC), pulsed (A,F), smooth DC (B,B+)
- Test currents (IΔN): .............10 mA, 15 mA, 30 mA;.................100 mA and 300 mA with A1322, A1422
- Test current size (PRCD standard is AS/NZS 3017)............... ±5%
- Test current size (general)………………-0/+10%
Test terminals: RCD Test Trip-out time (tΔN) - Range: 0 ms … 300 ms (999 ms*) (½xI∆N) / 0 ms … 300 ms (I∆N) / 0 ms … 40 ms (5xI∆N)
- Resolution: 1ms / 1ms / 1ms
- Accuracy: ±3ms / ±3ms / ±3ms
*According to standard AS/NZS 3017 Trip out-current (IΔ) - Range: 0.2xI∆N … 2.2xI∆N
- Resolution: 0.05xI∆N
- Accuracy: ±0.1xI∆N
Contact voltage (Uc) - Range: 0.0 V … 19.9 V / 20.0 V … 99.9 V
- Resolution: 0.1V / 0.1V
- Accuracy: (-0 % / +15 %) of reading ± 10 D / 0.1 V (-0 % / +15 %) of reading
- Test current type:.................sine-wave (AC), pulsed (A,F), smooth DC (B,B+)
- Test currents (IΔN): .............10 mA, 15 mA, 30 mA
- Test current size (RCD standard is AS/NZS AS/NZS)............... 5%
- Test current size (EN 61008/EN 61009)………………-0/+10%
Test terminals: PE conductor (PRCD) PE conductor (Type = 2 pole, 3 pole, S(3 pole)) (R) - Range: 0.00 Ω … 19.99 Ω / 20.0 Ω … 99.9 Ω / 100.0 Ω … 199.9 Ω / 200 Ω … 999 Ω
- Resolution: 0.01 Ω / 0.1 Ω / 0.1 Ω / 1 Ω
- Accuracy: ±(2 % of reading + 2 D) / ± 3 % of reading / ± 5 % of reading / indicative
- Operating range (acc. to EN 61557-4)............. 0.08 Ω … 199.9 Ω
- Current source (at nominal mains voltage, use of standard accessories).......................................................................... ..> 0.2 A at R < 2 Ω
- Open circuit voltage.......................................... ..< 9 V a.c.
Test terminals: PE conductor (Type = K/ Di (varistor)) Test principle: A voltage is applied between PE connections of the PRCD-K. There is a ‘PASS’ if PRCD trips. - Open circuit voltage..................... 24 V
- Output resistance......................... 220 Ω ± 10 % (IΔN = 30 mA), 620 Ω ± 10 % ((IΔN = 10 mA)
Test terminals: Open conductor (PRCD) Test principle: Mains voltage is applied to the mains test socket. Disconnection of the L, N and PE connections is performed inside the instrument. There is a ‘PASS’ if the PRCD trips. Test terminals: - L open, N open, PE open Socket – IEC
PRCD PE probe test Test principle: Mains voltage is applied to the mains test socket. A safe voltage sufficiently high to activate the protection circuit in the PRCD is applied to the P/S terminal. - Test voltage (active) ......................................... > 100 V a.c.
- Maximal current ................................................ < 1 mA
Test terminals: - Result (manual): Socket, P/S
- Result (auto): Socket – IEC, P/S
Polarity - Test voltage (normal).................................................................. < 50 V
- Test voltage (active)…………………………………………......... mains voltage
- Power consumption of tested device during the active test…….< 25 VA
Test terminals: - Result (normal, active) Socket – IEC
Clamp current (I, Idiff, Ipe) True RMS current using 1000:1 current clamp - Range: 0.10 mA … 9.99 mA / 10.0 mA … 99.9 mA / 100 mA … 999 mA / 1.00 A … 9.99 A / 10.0 A … 24.9 A
- Resolution: 0.01 mA / 0.1 mA / 1 mA / 0.01 A / 0.1 A
- Accuracy: ±(5 % of reading + 10 digits) / ±(5 % of reading + 5 digits) / ±(5 % of reading + 5 digits) / ±(5 % of reading + 5 digits) / ±(5 % of reading + 5 digits)
Accuracy of current transformer is not considered. Frequency range of current clamp is not considered. Test terminals: - I, Idiff, Ipe: CLAMPS terminals
Flash test Current a.c. (apparent) (I) - Range: 0.00 mA … 2.50 mA
- Resolution: 0.01mA
- Accuracy: ±(5 % of reading + 5 D)
- Open circuit test voltage..................... 1500 V, 3000 V (-0/+5%) at supply voltage 115 V, 230 V
- Short circuit current…………………….< 3.5 mA
- Output resistance: .................. ........... 480 kΩ@1500 V, 960 kΩ@3000 V
Test terminals: - I (1500 V): Socket LN – Socket PE
- I (3000 V): Socket LN – FLASH
Insulation resistance Riso (welding equipment) Riso - Range: 0.00 MΩ … 19.99 MΩ / 20.0 MΩ … 99.9 MΩ / 100.0 MΩ … 199.9 MΩ
- Resolution: 0.01 MΩ / 0.1 MΩ / 0.1 MΩ
- Accuracy: ±(3 % of reading + 2 D) / ±5 % of reading / ±10 % of reading
Output voltage (Um) - Range: 0 V … 600 V
- Resolution: 1 V
- Accuracy: ±(3 % of reading + 2 D)
- Operating range (acc. to EN 61557-2).........0.08 MΩ … 199.9 (999) MΩ
- Nominal voltages Un ....................................500 V (- 0 %, + 10 %)
- Short circuit current ......................................max. 2.0 mA
Test terminals: - Riso LN-W: A 1422: Socket L1 L2 L3 N (16A-5p, 32A-5p or 16A-3p) – W1 W2
- Riso W-PE: A 1422: Socket PE (16A-5p, 32A-5p or 16A-3p) – W1 W2
- Riso LN-PE: A 1422: Socket L1 L2 L3 N (16A-5p, 32A-5p or 16A-3p) – Socket PE (16A-5p, 32A-5p or 16A-3p)
- Riso LN (Class II) - P: A 1422: Socket L1 L2 L3 N (16A-5p, 32A-5p or 16A-3p) – MI 3360: P/S
Insulation Resistance (medical equipment) Riso - Range: 0.00 MΩ … 19.99 MΩ / 20.0 MΩ … 199.9 MΩ
- Resolution: 0.01 MΩ / 0.1 MΩ
- Accuracy: ±(3 % of reading + 2 D) / ±5 % of reading
Output voltage (Um) - Range: 0 V … 600 V
- Resolution: 1 V
- Accuracy: ±(3 % of reading + 2 D)
- Operating range (acc. to EN 61557-2).........0.08 M … 199.9 (999) M
- Nominal voltages Un ....................................500 V (- 0 %, + 10 %)
- Short circuit current ......................................max. 2.0 mA
Test terminals: - Riso (LN-PE) LN, Socket LN – PE, Socket PE
- Riso (LN-P/S) LN, Socket LN – P/S
- Riso (LN-AP) LN, Socket LN – P/AP
- Riso (PE-AP) PE, Socket PE – P/AP
Equipment leakage Equipment leakage current (direct, differential, alternative) (Ieq) - Range: 0.000 mA … 1.999 mA / 2.00 mA … 19.99 mA
- Resolution: 1 µA / 0.01 mA
- Accuracy: ±(3 % of reading + 3 D) / ±(5 % of reading)
Ulpe (direct, differential, alternative) - Range: 0 V … 299 V
- Resolution: 1 V
- Accuracy: ±(2 % of reading + 2 D)
Power (direct, differential) (P) - Range: 0 W … 999 W / 1.00 kW … 3.70 kW
- Resolution: 1W / 10W
- Accuracy: ±(5 % of reading + 5 D) / ±5 % of reading
- Operating range direct and differential method (acc. to EN 61557-16)…..0.010 mA … 19.99 mA
- Operating range alternative method (acc. to EN 61557-16)……………….0.020 mA … 19.99 mA
- Influence of load current (differential method) . …………….< 0.02 mA/A
Test terminals: - Ieq (alternative): LN, Socket LN – PE, Socket PE, P/S, P/AP
- Ieq (direct, differential): Socket L,N – PE, Socket PE, P/S, P/AP
- Ulpe: Socket L – Socket PE
- P (direct, differential): Socket L – Socket N
Applied Part leakage Applied Part leakage current (direct, alternative) (Iap) - Range: 0.000 mA … 1.999 mA / 2.00 mA … 19.99 mA
- Resolution: 1 µA / 0.01 mA
- Accuracy: ±(3 % of reading + 3 D) / ±(5 % of reading)
Uap (direct, alternative) - Range: 0 V … 299 V
- Resolution: 1 V
- Accuracy: ±(2 % of reading + 2 D)
Power (direct (P)) - Range: 0 W … 999 W / 1.00 kW … 3.70 kW
- Resolution: 1 W / 10 W
- Accuracy: ±(5 % of reading + 5 D) / ±5 % of reading
- Operating range direct method (acc. to EN 61557-16)…………0.010 mA … 19.99 mA
- Operating range alternative method (acc. to EN 61557-16)……0.020 mA … 19.99 mA
Test terminals: - Iap (alternative): Socket LNPE, PE – P/AP
- Iap (direct): Socket L,N,PE, PE – P/AP
- Uap: Socket PE, PE – P/AP
- P: Socket L – Socket N
Touch current (medical equipment) Touch current (Itou) - Range: 0.000 mA … 1.999 mA / 2.00 mA … 19.99 mA
- Resolution: 1 µA / 0.01 mA
- Accuracy: ±(3 % of reading + 3 D) / ±(5 % of reading)
Ulpe (direct) - Range: 0 V … 299 V
- Resolution: 1 V
- Accuracy: ±(2 % of reading + 2 D)
Power (direct) (P) - Range: 0 W … 999 W / 1.00 kW … 3.70 kW
- Resolution: 1 W / 10 W
- Accuracy: ±(5 % of reading + 5 D) / ± 5 % of reading
- Operating range (acc. to EN 61557-16)........... 0.010 mA … 19.99 mA
Test terminals - Itou: Socket L,N – P/S
- Ulpe: Socket L – Socket PE
- P: Socket L – Socket N
General data Mains supply - Supply voltage, frequency .......................110 V / 230 V AC, 50 Hz / 60 Hz
- Supply voltage tolerance .........................10 %
- Max. power consumption.........................300 VA (without load on test socket)
- Max. load………………………. ...............10 A continuous, 16 A short duration, 1.5 kW motor
- Mains supply overvoltage category.........CAT II / 300V
- Altitude.....................................................≤ 2000 m
Measuring categories - Instrument:...............................................Cat II / 300 V
- Test socket:..............................................Cat II / 300 V
- Plug test cable: ........................................Cat II / 300 V
- Altitude.....................................................≤ 2000 m
Protection classifications - Power supply……………………… ..........Class I, mains supply, Class II, only battery supply
- Pollution degree.......................................2
- Degree of protection ................................IP 40
- .................................................................IP 20 (mains test socket)
- Case.........................................................Shock proof plastic / portable
- Display .....................................................Colour TFT display, 4.3 inch, 480 x 272 pixels
- Touch screen ...........................................Capacitive
Communication - Memory....................................................depends on microSD card size
- RS232 interfaces .....................................2
- USB 2.0....................................................Standard USB Type B
- Bluetooth..................................................Class 2
- Dimensions (w×h×d):...............................31 cm × 13 cm × 25 cm
- Weight......................................................6.1 kg
Reference conditions - Reference temperature range: ................15 ˚C … 35 ˚C
- Reference humidity range: ......................35 % … 65 % RH
- Operation conditions
- Working temperature range:....................0 ˚C … +40 ˚C
- Maximum relative humidity:.....................85 % RH (0 ˚C … 40 ˚C), non-condensing
- Storage conditions
- Temperature range:.................................-10 ˚C … +60 ˚C
- Maximum relative humidity:.....................90 % RH (-10 ˚C … +40 ˚C)
- .................................................................80 % RH (40 ˚C … 60 ˚C)
Accuracies apply for 1 year in reference conditions. Temperature coefficient outside these limits is 0.2 % of measured value per ˚C plus 1 digit, otherwise noted. Fuses: 2 x T 16 A / 250 V, 32 mm 6.3 mm / 1500 A |