Rigel Medical SafeTest 60 PAT Tester

SKU: RIGELSAFETEST60
  • Performs numerous tests for comprehensive PAT testing even by a novice
  • High current, low energy bond testing ensures accuracy and precision
  • Specifically designed for PAT testing on medical and laboratory equipment
More Info
Typically ships in 7-10 working days

Rigel Medical SafeTest 60 PAT Tester

The Rigel Medical SafeTest 60 PAT Tester is an ideal instrument for ensuring the compliance of medical and laboratory equipment that does not require patient lead testing. Examples include hoists, ventilators, nebulisers, syringe drivers and much more.

The Rigel SafeTest 60 tests a wide range of electrical variables including current leakage, earth bond, touch bond, mains voltage, load current, point-to-point, and touch voltage as well as performing insulation testing. The high current and low energy bond testing guarantees accuracy and precision whilst still testing at 200mA. It has been specifically designed for the medical and scientific research sectors and can test to all pertinent international standards.

The test process has been made very simple thanks to automatic mains configuration verification and a one-button-one-function interface so even a novice can use the tester with confidence. The user can manually set fault conditions on the Rigel SafeTest 60 thereby saving time and performing only the tests required.

A step-through guide for each test routine helps eliminate errors and operation is conducted with the aid of an extremely clear colour screen. The screen clearly displays test results as well as the Rigel SafeTest 60’s intuitive menus.

Aside from comprehensive PAT testing functions and an intuitive interface, the Rigel SafeTest 60 is a compact, lightweight instrument. It can easily be transported thanks to an included carry case which has space for the tester itself and all accessories including a bond test probe, detachable mains cable and calibration certificate. A 2-year warranty is provided by the manufacturer.

Key Features

  • Designed for medical and laboratory uses
  • Current leakage, earth bond, touch bond, mains voltage
  • Load current, point-to-point, insulation testing, touch voltage
  • Accurate high current (200mA), low bond earth testing
  • Automatic mains configuration verification
  • Manual adjustment of fault conditions
  • Vibrant colour screen
  • Intuitive, novice-friendly operation
  • Two-year warranty

What's In The Box?

  • Rigel Medical SafeTest 60 PAT Tester
  • Calibration certificate
  • Carrying case
  • Bond test probe with clip
  • Detachable mains cable
  • Electronic instruction manual
  • Quick start guide
Specification

Earth Continuity

  • Method: 2 wire ZAP technique.
  • Test Current: >± 200mADC into 2Ω
  • Max Test Voltage: 4-24Vrms o/c
  • Measuring Range (low range): 0.001 – 0.999Ω / 0.001Ω resolution
  • Measuring Range (mid range): 1.00 – 9.99Ω / 0.01Ω resolution
  • Measuring Range (high range): 10.0 – 19.9Ω / 0.1Ω resolution
  • Accuracy: ± 1% of value, ± 5mΩ
  • Circuit Protection: Test inhibited if ≥30VAC or DC at 4mm inputs

Insulation Resistance

  • Measurement: EUT to Earth
  • Voltage: 500VDC @1mA.
  • Maximum O/C Voltage: <600VDC
  • Range: 100K - 20MΩ ± 5% ± 2 digits
  • Resolution: 0.01MΩ
  • Short circuit current: <2mA

Powered Leakage Measurements

  • IEC 62353: Equipment Leakage (Direct)
  • IEC / AAMI 60601: Earth + Enclosure Leakage
  • NFPA-99: Ground + Chassis Leakage
  • IEC 61010: Touch Leakage, Touch Voltage
  • Test Voltage: Mains Supply Voltage
  • Measuring Range: 0.1- 9999µA (0.1-8000µA typical for IEC 61010)
  • Measurement/Display Resolution: 0.1µA
  • Accuracy: ± 2%, ± 5µA
  • Mains Reversal: Softkey
  • Single Fault Conditions: Open neutral via softkey; Open earth via softkey
  • Frequency Response: IEC 60601 – 62353, NFPA-99 and IEC 61010

Voltage measurement

  • Application: L-N, L-E, N-E and touch voltage (IEC 61010)
  • Range: 0.0V – 300VAC
  • Resolution: 0.1V
  • Accuracy: ± 2% ± 2 digits (between 10V – 270VAC)
  • Mains frequency: 45.0 – 66.0Hz
  • Resolution: 0.1Hz
  • Accuracy: Unspecified

EUT Load Current Measurement

  • Range: 0.0A – 20.0A
  • Resolution: 0.1A
  • Accuracy: ±5% ± 2 digits

Power Source

  • Maximum current rating: 20A @ 120V / 16A @ 230V
  • Duty cycle: 15A to 20A, 5 min. on/5 min. off; 10A to 15A, 7 min. on/3 min. off; 0A to 10A, continuous
  • Mains power: 90-264V 48-64Hz
Label Type Size Download
pdf Rigel SafeTest 60 Datasheet pdf 1.2 MB Download