TPI 717R Flue Gas Analyser Kit 5

SKU: TPI717RKIT5
  • TPI 719 Gas Leak Detector for sensing dangerous organic gases
  • Measures temperature, pressure, carbon monoxide concentration and oxygen
  • CK21M pipe clamp probes make use of differential temperature function
More Info
Discontinued

TPI 717R Flue Gas Analyser Kit 5

The TPI 717R Flue Gas Analyser Kit 5 contains the adaptable TPI 717R alongside several useful accessories which include an A740 IR printer, two CK21M pipe clamp temperature probes, a CPA1 probe set for flue-less appliances, and a TPI 719 gas leak detector for hydrocarbon gas detection.

The analyser itself measures several different parameters including oxygen level, carbon monoxide level, temperature, and pressure. Differential pressure and temperature calculation on the TPI 717R allows advanced analysis of combustion system performance whilst the calculation system can provide values for CO2, CO/CO2 ratio, efficiency and excess air content. There are also options for timed tightness tests and timed ambient CO build-up tests. The CPA1 probe set allows measurement on flue-less installations.

Both the TPI 717R and 719 contribute to maintaining a safe working environment. The analyser can measure ambient CO levels and will not shut down when dangerous concentrations are detected whilst the gas leak detector can sense a wide range of organic gases and has extremely high sensitivity.

Results are shown on the analyser’s clear, backlit display which also gives a countdown to the next service due date when the TPI 717R is switched on. A maximum of 100 data sets can be saved to the internal memory and then wirelessly transferred to the infrared printer for printing on site.

Water and dust filters in the flue gas probe help prevent ingress into the analyser and it is also protected by tough rubber-boot cladding. With a six-year warranty, the TPI 717R is an analyser you can rely on and it is perfectly supplemented by the accessories contained in Kit 5.

TPI part number: 717R-Kit 5

Key Features

  • O2 range: 0 to 25%
  • CO range: 0 to 10000ppm
  • CO2 range: 0 to 25%
  • Temperature range: -40˚C to 800˚C
  • Pressure range: -150mBar to +150mBar
  • Differential pressure/temperature
  • Efficiency & excess air calculation
  • Tightness/let by testing
  • Ambient CO measurement
  • Gas leak detector
  • Wireless infrared printer
  • CPA1 probe set for flue-less appliances
  • 2 x CK21M pipe clamp temperature probes
  • 6-year warranty

We offer the following kits for the TPI 717R:

  Standard Kit Kit 1 Kit 2 Kit 3 Kit 4 Kit 5 Oil Kit
TPI 717R Flue Gas Analyser
A740 IR Printer  
2 x CM21K Temperature Pipe Clamp Probes      
CPA1 Probe Set          
TPI 719 Gas Leak Detector          
Oil filter            
Smoke Pump            

 

TPI 717R Kit Featured Contents

  • TPI 717R Flue Gas Analyser

TPI 717R Kit 1 Featured Contents

  • TPI 717R Flue Gas Analyser
  • TPI A740 IR Printer

TPI 717R Kit 2 Featured Contents

  • TPI 717R Flue Gas Analyser
  • TPI A740 IR Printer
  • 2 x CK21M Pipe Clamps

TPI 717R Kit 3 Featured Contents

  • TPI 717R Flue Gas Analyser
  • TPI A740 IR Printer
  • 2 x CK21M Pipe Clamps
  • CPA1 Probe Set

TPI 717R Kit 4 Featured Contents

  • TPI 717R Flue Gas Analyser
  • TPI A740 IR Printer
  • 2 x CK21M Pipe Clamps
  • TPI 719 Gas Leak Detector

TPI 717R Kit 5 Featured Contents

  • TPI 717R Flue Gas Analyser
  • TPI A740 IR Printer
  • 2 x CK21M Pipe Clamps
  • CPA1 Probe Set
  • 719 Gas Leak Detector

TPI 717R Oil Kit Featured Contents

  • TPI 717R Flue Gas Analyser
  • TPI A740 IR Printer
  • Smoke pump
  • Oil filter

What's In The Box?

  • TPI 717R Flue Gas Analyser
  • Flue Temperature Probe
  • ‘K’-Type Temperature Air Probe GK11M
  • In-Line Water Trap Bowl Filter Complete
  • Mini Pump Protection Filter Assembly
  • Battery Charger
  • Protective Boot
  • Soft Carrying Case
  • Exhaust Spigot
  • TPI A740 IR Printer
  • 2 x TPI CK21M Pipe Clamp Temperature Probes
  • TPI 719 Gas Leak Detector
  • CPA1 probe set
Specification

Carbon Monoxide (CO) Measurement

  • Range: 0 to 10,000 ppm
  • Resolution: 1 ppm
  • Accuracy: <20 ppm : ± 3ppm; >100 ppm : ± 5%

Oxygen (O2) Measurement

  • Range: 0 to 25%
  • Resolution: 0.1%
  • Accuracy: ±0.3% 

Carbon Dioxide (CO2)

  • Range: 0 to 25%
  • Resolution: 0.1%
  • Accuracy: Calculated

Ratio (CO/CO2)

  • Range: 0 to 0.999
  • Resolution: 0.001
  • Accuracy: Calculated 

Gross/Net Efficiency

  • Range: 0 to 100%
  • Resolution: 0.1%

Temperature Measurement

  • Range: -40° to 800° C (-40° to 1472° F)
  • Resolution: 1°C or F
  • Accuracy (K-type): ±0.3%, ±1°C or ±2°F

Draft/Pressure Measurement

  • Selectable Ranges: mBar,kPa, and inH2O
  • Range: -150 mBar to +150 mBar
  • Resolution: 0.01mBar
  • Accuracy: ±0/5% fsd 

Instrument Specifications

  • Operating Temperature Range: -10° to 50°C
  • Battery: Rechargeable Ni-MH, >6 Hour Life
  • Charger Input Voltage: 115V or 230V; 50/60 Hz AC
  • Fuels: Natural Gas, LPG, Light Oil, Heavy Oil, Bituminous Coal, Anthracite Coal, Coke, Butane, Wood (Dry), Bagasse, Wood Pellet
  • Display Backlit: LCD
  • Data Storage: 100 sets
  • Time & Date: 24 Hour Clock
  • Dimensions: 200 x 90 x 60mm
  • Weight: 500g
  • Conforms To: EN50379 / BS7967

TPI 719 Combustible Gas Detector

  • Gases detected: Methane, Natural Gas, Gasoline, Diesel, Propane (LPG), Butane, Kerosene
  • Sensitivity: 30ppm (CH4)
  • Sensor Type: Low power semiconductor
  • Probe Length: 16”
  • Leak Indication: Audio / Visual
  • Adjustable Tic Rate: Yes
  • Battery: “C” cell (2)
  • Weight: 500g
Label Type Size Download
pdf TPI 717R Datasheet pdf 414.1 KB Download