Specification
Technical Specifications
Fluke 718 Series
- Measure pressure - (Internal sensor)
- Accuracy - 0.025%, full scale
- Functions - Zero, min/max, hold, damp
- DRY AIR, NON-CORROSIVES ONLY
Fluke 718 1G
- Range - -1psi to + 1psi, -68.9mbar to 68.9mbar (-6.89kPa to 6.89kPa)
- Resolution - 0.0001psi, 0.001mbar, 0.0001kPa
- Over pressure - 5x full scale
- Functions - Zero, Min, Max, Hold, Damp
Fluke 718 30G
- Range - -12psi to 30psi, (-850mbar to 2bar, -85 to 206.84kPa)
- Resolution - 0.001psi, 0.1mbar, 0.01kPa
- Over pressure - 5x full scale
- Functions - Zero, Min, Max, Hold, Damp
Fluke 718 100G
- Range - -12psi to 100psi, (-850mbar to 7bar, -85 to 689.48kPa)
- Resolution - 0.01psi, 1mbar, 0.1 kPa
- Over pressure - 2x full scale
- Functions - Zero, Min, Max, Hold, Damp
Fluke 718 300G
- Range - -12psi to 300psi, (-850mbar to 20bar, -85 to 2068.42kPa)
- Resolution - 0.01psi, 1mbar, 0.1kPa
- Over pressure - 375psi, 25bar
- Functions - Zero, Min, Max, Hold, Damp
Fluke 718 Series via 700 Series Pressure Modules
- Range - 29 pressure modules, 0 – 1 inH20 to 10,000psi, 2.5mbar to 700bar. Over pressure per pressure module specs
- Resolution - Per pressure module specs
- Accuracy - To 0.025% of full span, per pressure module specs
- Functions - Zero, Min, Max, Hold, Damp
- MEDIA COMPATIBILITY PER PRESSURE MODULE SPECS
Fluke 718 Series via Built in Pump
- Range - -12psi or -850mbar to full scale Supported Pressure Units; psi, inH20 (4°C), inH20 (20°C), cmH20 (4°C), cmH20 (20°C), bar, mbar, kPa, inHg, mmHg, kg/cm
- Loop power range - 24 V DC
- Accuracy - ±10%
- Drive; 20mA into 1000 Ω for battery > 6.8V; 700 Ω for battery 5.8 to 6.8V
mA Measure
- Range - 0mA to 24mA
- Resolution - 0.001mA
- Accuracy - 0.015% + 1 count
General Specifications
- Maximum Voltage - 30 volts (non-operating)
- Temperature - -40°C to 60°C
- Operating temperature - -10°C to 55°C
- Relative humidity - 95% (10°C to 30°C); 75% (30°C to 40°C); 45% (40°C to 50°C); 35% (50°C to 55°C)
- Operating altitude - 3,000 M max
- Shock - 1M drop tested
- Vibration - Random, 2g, 5Hz - 500Hz
- Safety - CSA C22.2 No. 1010.1: 1992
- EMC - EN50082-1: 1992 and EN55022: 1994 Class B
- Size - 216 x 94 x 66mm (8.50 x 3.72 x 2.60in)
- Weight - 992g (35oz)
- Power, battery life - 9V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline; two batteries in 718
- Battery life - 4 to 20 hours typical, depending on functions used
- Warranty - Three years (one year on pressure pump in Fluke 718)
- Display - LCD, 5 digit pressure and current simultaneous
- For use with non-corrosive gasses
|