Fluke 729 Automatic Pressure Calibrator

SKU: FLUKE729-CONF
  • Automatic pressure calibrator for generating desired pressure with an accuracy of 0.01% ±2 LSD all ranges (at 23 °C ±5 °C)
  • Can be purchased with either 30 psi, 150 psi, or 300 psi, representing a wide scope for pressure calibration
  • IP54 rated which protects against dust and water, making it suitable for use in a range of environments
More Info

"Buy a Fluke, get a FREE Fluke" promotion. Spend £150 or more on any qualified Fluke product(s) and claim a FREE Fluke gift!
[Click here to learn more] Offer is valid from 1st March to 31st August 2024.

Fluke 729 Automatic Pressure Calibrator

The Fluke 729 Automatic Pressure Calibrator has been expertly designed to provide faster, more accurate test results, and reduce complications within the process of pressure calibration. With an internal electric pump that provides automatic pressure generation and regulation, the 729 makes the job easier for process technicians by combining precision results with a simplified design. Simply type in a target pressure up to 300psi, and the 729 Pressure Calibrator will automatically pump up to and stabilise at the requested value. The user can also fill in a test template which will pump to and document a multi-point pressure calibration test at the touch of a button.

The 729 (without FlukeConnect and wireless connection capabilities) range consists of three different PSI-level calbibrator's, which are the 729 30G, THE 729 150G, and the 729 300G. These calibrators can test multiple pressure points and document the results aswell. Simply let the 729 do the work, as it documents and flags-out applied pressure, measured mA, mV, digital PV and the percentage error for each test point. The user can choose from three ranges: 30 psi (2 bar, 200 kPa), 150 psi (10 bar, 1MPa), and 300 psi (20 bar, 2MPa).

HART Communication

Built in HART communication capabilities allow for HART transmitter mA adjustments, HART configuration, and the ability to adjust to applied 0% and 100% values. Aswell as these abilities, the 729 is also able to set fixed mA outputs for troubleshooting, variables & read device diagnostics, and read device configuration.

Convenient and Ergonomic Design

The 729 has been one-meter drop tested and is able to withstand any accidents that may occur on the work site, making it a great choice for industry professionals who are looking to calibrate pressure. The 729 is also supplied with a standard semi-rigid carrying case, which is designed to withstand all types of field work, with room to store test hoses, leads and small tools which can be used to suit whatever situation is at hand.

The Fluke 729 30G has a psi range of -12.0000 to 30.0000, the 150G has one of -12.000 to 150.000 psi, and the 300G has a psi of -12.000 to 300.000, meaning there is room to select the perfect product for the job at hand. To compare the three models, read our comparison table and decide which one suits your needs. Please note: due to the three models listed above not having wireless capabilities, they cannot be connected to Fluke Connect. For this ability, explore the Fluke 729FC range.

Fluke part number: FLK-729 30G, FLK-729 150G, FLK-729 300G

Key Features

  • Offers automatic pressure calibration
  • Operating temperature of -10°C to 50°C for measurement, 0°C to 50°C for pressure control
  • Up to 300psi (with selected model)
  • Built in HART communication for sending and receiving digital information
  • 1 meter drop tested
  • IP54 rated for protection against dust, water, etc.
  • Accuracy of 0.01% ±2 LSD all ranges (at 23 °C ±5 °C)

Fluke 729 (Without FC) Comparison Table

  729 30G 729 150G 729 300G
Psi Range, Resolution -12.0000 to 30.0000 psi -12.000 to 150.000 psi -12.000 to 300.000 psi
Bar Range, Resolution -0.82737 to 2.06842 bar -0.8273 to 10.3421 bar -0.8273 to 20.6843 bar
kPa Range, Resolution -82.737 to 206.843 kPa -82.73 to 1034.21 kPa -82.73 to 2068.43 kPa
Comment No wireless communication

What's In The Box?

  • Fluke 729 Automatic Pressure Calibrator
  • Li-Ion battery charger and adaptors
  • Quick-Connect pressure fitting, hose and adaptor starter kit
  • 1 pair of stackable test leads
  • 1 set of AC280 SureGrip hook clips
  • 1 set of AC175 alligator clips
  • 1 pair of TP220 test probes (red, black)
  • USB communication cable
  • TPAK ToolPak Magnetic Meter Hanger Carrying case with shoulder strap and accessories pouch
  • DPCTrack2 trial calibration management software
  • Traceable certificate of calibration
  • Quick Reference Guide
  • 3-year warranty
Specification

Ranges

  • mA Measuer, Source and Simulate - 0 mA to 24 mA
  • Volts DC Measure - 0 V dc to 30 V dc

Resolution

  • mA DC Source, Simulate and Measure - 1 μA
  • Voltage DC Measurement - 1 mV
  • Accuracy - 0.01% ±2 LSD all ranges (at 23°C ±5°C)
  • Stability - 20ppm of full scale/°C from -10C + 18°C and 28°C to + 50°C
  • mA Simulate External Voltage Requirement - 12 V dc to 30 V dc
  • Loop Compliance Voltage - 24 V dc @ 20 mA
  • mA Simulate External Voltage Requirement - 12 V dc to 30 V dc
  • Temperature Measurement only/100 Ω Pt(385) RTD
  • Temperature Resolution - 0.01°C (0.01°F)
  • Temperature Accuracy - ±0.1°C (0.2°F) ±0.25°C (±0.45°F) combined uncertainty when using 720 RTD probe (optional accessory)
  • Drive Capability - 1200 Ω without HART resistor, 950 Ω with internal HART resistor

Mechanical Specification

  • Size (HxWxL) - 7 x 27.94 x 17.27cm (2.75 x 11 x 6.8in)
  • Weight - 6.5lbs, 2.95kg
  • Ingress Protection - IP54

Environmental Specification

  • Operating Temperature - -10°C to 50°C for measurement, 0°C to 50°C for pressure control
  • Operating Temperature with Battery - -10°C to 40°C (Battery will only charge to 40°C)
  • Storage Temperature - -40°C to 60°C
  • Operating Altitude - 3000m
  • Storage Altitude - 13000m