Fluke 753 Documenting Process Calibrator

SKU: FLUKE753
  • Measure volts, mA, RTDs, thermocouples, frequency and ohms for testing sensors and transmitters
  • Power a transmitter during a test using loop supply with simultaneous mA measurement
  • Measure Voltage, current, resistance, frequency, temperature, pressure and Store up to a weeks worth of procedures and calibration results
More Info
Typically ships in 7-10 working days

"Buy a Fluke, get a FREE Fluke" promotion. Spend £150 or more on any qualified Fluke product(s) and claim a FREE Fluke gift!
[Click here to learn more] Offer is valid from 1st March to 31st August 2024.

Fluke 753 Documenting Process Calibrator

The Fluke 753 Documenting Process Calibrator is a multifunctional calibrator allowing users to download procedures, lists and instructions and measure a range of parameters such as volts, mA, RTDs, thermocouples, frequency and ohms for testing sensors and transmitters. The Fluke 753 can power a transmitter during a test busing the loop supply with simultaneous mA measurements.

The Fluke 753 sources, simulates and measures pressure, temperature and electrical signals in one rugged hand held calibration device. Automated calibration procedures and data documentation can be captured with the Fluke 753 meeting industry standards such as ISO 9000, FDA, EPA and OSHA regulations

With a USB interface and USB communication cable, the Fluke 753 enables two way Communication with DPCTrack 2 and other instrument management applications/ Using the Fluke DPCTrack2 Calibration Management software to create a seamless and paperless calibration management system.

Users are able to store up to a weeks worth of downloaded procedures and calibration results on the device, the Fluke 753 offers a number of flexible feature such as autostep, custom units, user-entered values during a test, one point and two point switch testing, square root DP flow testing and programmable measurements delay.

The Fluke 753 boasts a bright white dual display allowing for reading both sourced and measured parameters simultaneously. Battery powered, the Fluke 753 has a battery life of around 8 hours allowing for a full days use in the field. IP52 rated and with a three year warranty, the Fluke 753 offers a robust and reliable device.

Key Features

  • Measure and sources pressure when used alongside Fluke 700Pxx Pressure Modules(sold separately)
  • Create and run automated as-found/as left procedures
  • Store up to a weeks work of calibration results
  • Measure Voltage, current, resistance, frequency, temperature, pressure
  • Source/simulate volts, mA, thermocouples, RTDs, frequency, Ohms and pressure
  • Power a transmitter during loop supply tests with simultaneous mA measurements
  • Rechargeable batteries allowing up to eight hours use on a full charge
  • IP 52 safety rated
  • Bright white dual display allowing for reading both sourced and measured parameters simultaneously
  • USB interface and USB communication cable, the Fluke 753 enables two way Communication with DPCTrack 2

What's In The Box?

  • Fluke 753 Documenting Process Calibrator
  • BC7240 battery charger
  • Li-on BP7240 battery pack
  • NIST-traceable calibration report and data
  • TP220 test probes x 3
  • "extended tooth" alligator clips x3
  • AC280 hook clips
  • C799 Soft Field Case
  • USB communication cable
  • Quick Reference Guide
  • 3 year warranty
Specification

Measurement Accuracy

Voltage DC

  • Range/Resolution100.000mV
    1 Year: 0.02% + 0.005mV
    2 Years 0.03% + 0.005 mV
  • Range/Resolution: 3.0000V
    1 Year:
    2 Years: 0.03% + 0.00005V
  • Range/Resolution: 30.0000V
    1 year: 0.02% + 0.0005V
    2 Years: 0.03% + 0.0005
  • Range/Resolution: 300.00V
    Year 1: 0.05% + 0.05V
    2years: 0.07% + 0.05V

Voltage AC

  • Range/Resolution3.000V (40Hz to 500Hz)/0.001V
    1 Year 0.5% + 0.002V
    2 Years 1.0% +0.004V
  • Range/Resolution30.000V (40Hz to 500Hz)/0.01V
    1 Year 0.5% + 0.00V
    2 Years 1.0% +0.04V
  • Range/Resolution300.0V (40Hz to 500Hz)/0.1V
    1 Year 0.5% + 0.2V
    2 Years 1.0% +0.2V

Current DC

  • Range/Resolution30.000mA
    1 Year 0.01% + 5uA
    2 Years 0.015% +7uA
  • Range/Resolution110.00mA
    1 Year: 0.01% +20uA
    2 Years: 0.015% +30uA

Resistance

  • Range/Resolution: 10.000 Ω
    1 Year: 0.05% + 50 mΩ
    2 Years: 0.07% + 70 mΩ
  • Range/Resolution: 100.00 Ω
    1 Year: 0.05% + 50 mΩ
    2 Years: 0.07% + 70 mΩ
  • Range/Resolution: 1.0000kΩ
    1 Year: 0.05% + 500 mΩ
    2 Years: 0.07% + 0.5Ω
  • Range/Resolution: 10.000kΩ
    1 Year: 0.1% + 10Ω
    2 Years: 0.15% + 15Ω

Frequency

  • Range/Resolution: 1.00 to 110.00Hz/0.01Hz
    Year 2 0.05Hz
  • Range/Resolution: 110.1 to 1100.00Hz/0.01Hz
    Year 2 0.05Hz
  • Range/Resolution: 1.101 to 11.000kHz/0.01Hz
    Year 2 0.005kHz
  • Range/Resolution: 11.01 to 50.00kHz/0.01Hz
    Year 2 0.05kHz

Source Accuracy

Voltage DC

  • Range/Resolution100.000mV
    1 Year: 0.01% + 0.005mV
    2 Years 0.015% + 0.005mV
  • Range/Resolution: 1.0000V
    1 Year: 0.01% +0.00005V
    2 Years: 0.015% + 0.00005V
  • Range/Resolution: 15.0000V
    1 year: 0.01% + 0.0005V
    2 Years 0.015% +0.0005V

Current DC

  • Range/Resolution 22.000mA (Source)
    1 Year 0.01% + 0.003 mA
    2 Years 0.02% + 0.003mA
  • Range/Resolution Current Sink (simulate)
    1 Year: 0.01% + 0.007mA
    2 Years: 0.04% + 0.007mA

Resistance

  • Range/Resolution: 10.000 Ω
    1 Year: 0.01% + 10mΩ
    2 Years: 0.015% + 15mΩ
  • Range/Resolution: 100.00 Ω
    1 Year: 0.01% + 20mΩ
    2 Years: 0.015% + 30mΩ
  • Range/Resolution: 1.0000kΩ
    1 Year: 0.02% + 0.2Ω
    2 Years: 0.03% + 0.3Ω
  • Range/Resolution: 10.000kΩ
    1 Year: 0.02% + 3Ω
    2 Years: 0.103% + 5Ω

Frequency

  • Range/Resolution: 0.1 to 10.99Hz
    Year 2 0.01Hz
  • Range/Resolution: 0.01 to 10.99Hz
    Year 2 0.01Hz
  • Range/Resolution: 11.00 to 109.99kHz
    Year 2 0.1Hz
  • Range/Resolution: 110.00 to 1099.9Hz/0.01Hz
    Year 2 0.1Hz
  • Range/Resolution: 1.100 to 21.999kHz
    Year 2 0.002kHz
  • Range/Resolution: 22.000 to 50.000kHz
    Year 2 0.005kHz

Technical Data

Data log functions

  • Measure Functions: Voltage, current, resistance, frequency, temperature and pressure
  • Reading Rate: 1, 2, 5, 10, 20, 30 or 60 readings/minute
  • Maximum record length: 8000 readings (7980 for 30 or 60 readings/minutes)

Ramp Functions

  • Source functions: Voltage, Current, resistance, frequency and temperature
  • Rate: 4 step/second
  • Trip detect: Continuity or voltage (continuity detection not available when sourcing current)

Loop power function

  • Voltage: selectable 26V
  • Accuracy: 10% 18V minimum at 22mA
  • Maximum current: 25mA short circuit protected
  • Maximum input voltage: 50V DC

Step Functions

  • Source functions: Voltage, current, resistance, frequency and temperature
  • Manual step: selectable step, change with arrow buttons
  • Autostep: Fully programmable for function, start delay, step value, time per step, repeat

Environmental specifications

  • Operating temperature: -10ºC to +50ºC
  • Storage temperature: -20ºC to 60ºC
  • Dust/Water resistance: Meets IP52, IEC 529
  • Operating altitude: 3000m above mean sea level (9842ft)

Safety specifications

  • Agency approvals: CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993

Mechanical and general specifications

  • Size: 136 x 245 x 63mm (5.4 x 9.6 x 2.5in)
  • Weight: 1.2KG (2.7lb(
  • Batteries: Internal battery pack Li-ion: 7.2V, 4400mAH, 30Wh
  • Battery life: >8 hours typical
  • Battery replacement: Via snap-shut door without opening calibrator, no tools required
  • Side port connections:
    Pressure module connector
    USB Connector to interface to your PC
    Connection for optional battery charger/eliminator
  • Data storage capacity: 1 week of calibration procedure results
  • 90 day specifications:
    The standard specification interval for the 750 series are 1 and 2 years
    Typical 90 day measurement and source accuracy can be estimated by dividing the one year "% of reading" or "% of output" specifications by 2
  • Floor specifications, expressed as "% of full scale" or "counts" or "ohms" remain constant

Temperature, Resistance Temperature Detectors

  • Degrees or % of reading - Type (a): 100Ω Pt (385)
    Range ºC: 1 Year -200 to 100 - 0.07ºC/100 to 800 - 0.02% + 0.05ºC
    2 Years: -200 to 100 - 0.14ºC/100 to 800 - 0.04% + 0.10ºC
  • Degrees or % of reading - Type (a): 200Ω Pt (385)
    Range ºC: 1 Year -200 to 100 - 0.07ºC/100 to 630 - 0.02% + 0.05ºC
    2 Years: -200 to 100 - 0.14ºC/100 to 630 - 0.04% + 0.10ºC
  • Degrees or % of reading - Type (a): 500Ω Pt (385)
    Range ºC: 1 Year -200 to 100 - 0.07ºC/100 to 630 - 0.02% + 0.05ºC
    2 Years: -200 to 100 - 0.14ºC/100 to 630 - 0.04% + 0.10ºC
  • Degrees or % of reading - Type (a): 1000Ω Pt (385)
    Range ºC: 1 Year -200 to 100 - 0.07ºC/100 to 630 - 0.02% + 0.05ºC
    2 Years: -200 to 100 - 0.14ºC/100 to 630 - 0.04% + 0.10ºC
  • Degrees or % of reading - Type (a): 100Ω Pt (3916)
    Range ºC: 1 Year -200 to 100 - 0.07ºC/100 to 630 - 0.02% + 0.05ºC
    2 Years: -200 to 100 - 0.14ºC/100 to 630 - 0.04% + 0.10ºC
  • Degrees or % of reading - Type (a): 100Ω Pt (3926)
    Range ºC: 1 Year -200 to 100 - 0.08ºC/100 to 630 - 0.02% + 0.06ºC
    2 Years: -200 to 100 - 0.16ºC/100 to 630 - 0.04% + 0.12ºC
  • Degrees or % of reading - Type (a): 10Ω Cu (427)
    Range ºC: 1 Year -100 to 260 - 0.2ºC
    2 Years: -100 to 260 - 0.4ºC
  • Degrees or % of reading - Type (a): 12Ω Ni (672)
    Range ºC: 1 Year -80 to 260 - 0.1ºC
    2 Years: -100 to 260 - 0.2ºC
  • Source Current: 1mA
    Source ºC 1 Year: 0.05ºC/ 0.0125% + 0.04ºC
    2 Years 0.10ºC/0.025% + 0.08ºC
    Allowable current: 0.1mA to 10mA
  • Source Current: 500 μA
    Source ºC 1 Year: 0.06ºC/ 0.017% + 0.05ºC
    2 Years 0.12ºC/0.034% + 0.10ºC
    Allowable current: 0.1mA to 10mA
  • Source Current: 250μA
    Source ºC 1 Year: 0.06ºC/ 0.017% + 0.05ºC
    2 Years 0.12ºC/0.034% + 0.10ºC
    Allowable current: 0.1mA to 10mA
  • Source Current: 150μA
    Source ºC 1 Year: 0.06ºC/ 0.017% + 0.04ºC
    2 Years 0.10ºC/0.025% + 0.08ºC
    Allowable current: 0.1mA to 10mA
  • Source Current: 1mA
    Source ºC 1 Year: 0.05ºC/ 0.0125% + 0.04ºC
    2 Years 0.10ºC/0.025% + 0.08ºC
    Allowable current: 0.1mA to 10mA
  • Source Current: 1mA
    Source ºC 1 Year: 0.05ºC/ 0.0125% + 0.04ºC
    2 Years 0.10ºC/0.025% + 0.08ºC
    Allowable current: 0.1mA to 10mA
  • Source Current: 3mA
    Source ºC 1 Year: 0.2ºC
    2 Years 0.08ºC
    Allowable current: 0.1mA to 10mA
  •  
  • Source Current: 1mA
    Source ºC 1 Year: 0.04ºC
    2 Years 0.08ºC
    Allowable current: 0.1mA to 10mA

Temperature, Resistance Temperature Detectors

  • Type: E
    Source ºC: -250º to -200/Measure ºC 1 year: 1.3/ 2 Years: 2.0/ Source ºC 1 year: 0.6/ 2 Years 0.9
    Source ºC: -200º to -100/Measure ºC 1 year: 0.5/ 2 Years: 0.8/ Source ºC 1 year: 0.3/ 2 Years 0.4
    Source ºC: -100º to 600/Measure ºC 1 year: 0.3/ 2 Years: 0.4/ Source ºC 1 year: 0.3/ 2 Years 0.4
    Source ºC: 600º to 1000/Measure ºC 1 year: 0.4/ 2 Years: 0.6/ Source ºC 1 year: 0.2/ 2 Years 0.3
  • Type: N
    Source ºC: -200º to -100/Measure ºC 1 year: 1.0/ 2 Years: 1.5/ Source ºC 1 year: 0.6/ 2 Years 0.9
    Source ºC: -100º to 900/Measure ºC 1 year: 0.5/ 2 Years: 0.8/ Source ºC 1 year: 0.5/ 2 Years 0.8
    Source ºC: 900º to 1300/Measure ºC 1 year: 0.6/ 2 Years: 0.9/ Source ºC 1 year: 0.3/ 2 Years 0.4
  • Type: J
    Source ºC: -210º to -100/Measure ºC 1 year: 0.6/ 2 Years: 0.9/ Source ºC 1 year: 0.3/ 2 Years 0.4
    Source ºC: -100º to 800/Measure ºC 1 year: 0.3/ 2 Years: 0.4/ Source ºC 1 year: 0.2/ 2 Years 0.3
    Source ºC: 800º to 1200/Measure ºC 1 year: 0.5/ 2 Years: 0.8/ Source ºC 1 year: 0.3/ 2 Years 0.3
  • Type: K
    Source ºC: -200º to -100/Measure ºC 1 year: 0.7/ 2 Years: 1.0/ Source ºC 1 year: 0.4/ 2 Years 0.6
    Source ºC: -100º to 400/Measure ºC 1 year: 0.3/ 2 Years: 0.4/ Source ºC 1 year: 0.3/ 2 Years 0.4
    Source ºC: 400 to 1200/Measure ºC 1 year: 0.5/ 2 Years: 0.8/ Source ºC 1 year: 0.3/ 2 Years 0.4
    Source ºC: 1200º to 1372/Measure ºC 1 year: 0.7/ 2 Years: 1.0/ Source ºC 1 year: 0.3/ 2 Years 0.4
  • Type: T
    Source ºC: -250º to -200/Measure ºC 1 year: 1.7/ 2 Years: 2.5/ Source ºC 1 year: 0.9/ 2 Years 1.4
    Source ºC: -200º to 0/Measure ºC 1 year: 0.6/ 2 Years: 0.9/ Source ºC 1 year: 0.4/ 2 Years 0.6
    Source ºC: 0 to 400/Measure ºC 1 year: 0.3/ 2 Years: 0.8/ Source ºC 1 year: 0.3/ 2 Years 0.4
  • Type: B
    Source ºC: 600º to 800/Measure ºC 1 year: 1.3/ 2 Years: 2.0/ Source ºC 1 year: 1.0/ 2 Years 1.5
    Source ºC: 800º to 1000/Measure ºC 1 year: 1.0/ 2 Years: 1.5/ Source ºC 1 year: 0.8/ 2 Years 1.2
    Source ºC: 1000 to 1820/Measure ºC 1 year: 0.9/ 2 Years: 1.3/ Source ºC 1 year: 0.8/ 2 Years 1.2
  • Type: R
    Source ºC: -20º to 0/Measure ºC 1 year: 2.3/ 2 Years: 2.8/ Source ºC 1 year: 1.2/ 2 Years 1.8
    Source ºC: 0º to 100/Measure ºC 1 year: 1.5/ 2 Years: 2.2/ Source ºC 1 year: 1.1/ 2 Years 1.7
    Source ºC: 100 to 1767/Measure ºC 1 year: 1.0/ 2 Years: 1.5/ Source ºC 1 year: 0.9/ 2 Years 1.4
  • Type: S
    Source ºC: -20º to 0/Measure ºC 1 year: 2.3/ 2 Years: 2.8/ Source ºC 1 year: 1.2/ 2 Years 1.8
    Source ºC: 0º to 200/Measure ºC 1 year: 1.5/ 2 Years: 2.1/ Source ºC 1 year: 1.1/ 2 Years 1.7
    Source ºC: 200 to 1400/Measure ºC 1 year: 0.9/ 2 Years: 1.4/ Source ºC 1 year: 0.9/ 2 Years 1.4
    Source ºC: 1400 to 1767/Measure ºC 1 year: 1.1/ 2 Years: 1.7/ Source ºC 1 year: 1.0/ 2 Years 1.5
  • Type: C
    Source ºC: 0º to 800/Measure ºC 1 year: 0.6/ 2 Years: 0.9/ Source ºC 1 year: 0.6/ 2 Years 0.9
    Source ºC: 800º to 1200/Measure ºC 1 year: 0.8/ 2 Years: 1.2/ Source ºC 1 year: 1.1/ 2 Years 1.7
    Source ºC: 200 to 1400/Measure ºC 1 year: 0.9/ 2 Years: 1.4/ Source ºC 1 year: 0.9/ 2 Years 1.4
    Source ºC: 1400 to 1767/Measure ºC 1 year: 1.1/ 2 Years: 1.7/ Source ºC 1 year: 1.0/ 2 Years 1.5
  • Type: L
    Source ºC: -200º to -100/Measure ºC 1 year: 0.6/ 2 Years: 0.9/ Source ºC 1 year: 0.3/ 2 Years 0.4
    Source ºC: -100 to 800º/Measure ºC 1 year: 0.3/ 2 Years: 0.4/ Source ºC 1 year: 0.2/ 2 Years 0.3
    Source ºC: 800 to 900/Measure ºC 1 year: 0.5/ 2 Years: 0.8/ Source ºC 1 year: 0.2/ 2 Years 0.3
  • Type: U
    Source ºC: -200º to 0/Measure ºC 1 year: 0.6/ 2 Years: 0.9/ Source ºC 1 year: 0.4/ 2 Years 0.6
    Source ºC: 0 to 600º/Measure ºC 1 year: 0.3/ 2 Years: 0.4/ Source ºC 1 year: 0.3/ 2 Years 0.4
  • Type: BP
    Source ºC: 0º to 1000/Measure ºC 1 year: 1.0/ 2 Years: 1.5/ Source ºC 1 year: 0.4/ 2 Years 0.6
    Source ºC: 1000 to 2000º/Measure ºC 1 year: 1.6/ 2 Years: 2.4/ Source ºC 1 year: 0.6/ 2 Years 0.9
    Source ºC: 2000 to 2500/Measure ºC 1 year: 2.0/ 2 Years: 3.0/ Source ºC 1 year: 0.8/ 2 Years 1.2
  • Type: XK
    Source ºC: -200º to 300/Measure ºC 1 year: 0.2/2 Years: 0.3/ Source ºC 1 year: 0.2 2 Years 0.5
    Source ºC: 300 to 800º/Measure ºC 1 year: 0.4/ 2 Years: 0.6/ Source ºC 1 year: 0.3/ 2 Years 0.6
Label Type Size Download
pdf Fluke 753 Documenting Process Calibrator Datasheet pdf 201.8 KB Download