Fluke 787B ProcessMeter™

The Fluke 787B cleverly amalgamates the functions of both a digital multimeter and a mA loop calibrator, thereby permitting comprehensive electrical inspections alongside accurate loop calibration. It can measure current up to 1A, voltage up to 1000V, resistance up 40MΩ, frequency up to 20kHz and also performs continuity and diode testing. True RMS measurement makes AC readings more accurate and the 787 boasts min/max/avg recording as well as display hold. To save time but still provide flexibility, the user can choose between automatic and manual ranging whilst relative mode permits compensation for test lead resistance amongst other uses.

From a loop calibration perspective, the Fluke 787B can source and simulate current between 4 and 20mA with an accuracy of ±0.05% and a precision of 0.001mA so it is suitable for most loop calibration functions. Different sourcing options are available with automatic slow ramp, fast ramp, manual coarse/fine/25% step and auto step. Readings are given in both milliamps and as a percentage for a comprehensive picture of loop performance.

Results for electrical measurement and loop calibration are shown on the Fluke 787B's clear, backlit screen. Operating the instrument is very easy despite the many different functions because of the clear rotating dial and simple on-screen symbols. The Cat III safety rating up to 1000V ensures that the user can have confidence in the device and it comes supplied with test leads, crocodile clips, test probes, and a rubber holster with an integrated stand. Overall, the Fluke 787B neatly combines accurate loop calibration capability with handy electrical measurement functions.

Fluke part number: 787B

Key Features

  • AC/DC voltage measurement: 0 to 1000V
  • AC/DC current measurement: 0 to 1A (440mA continuous)
  • Resistance measurement: 0 to 40MΩ
  • Frequency measurement: 0 to 20kHz
  • Continuity and diode testing
  • True RMS measurement
  • Min/max/average recording
  • Auto and manual ranging
  • Relative zero mode
  • Loop calibration sourcing and simulating
  • Test current from 4 to 20mA (0.05% accuracy, 0.001mA resolution)
  • Simple dial operation
  • Cat III to 1000V
  • Clear, backlit screen

What's In The Box?

  • Fluke 787B Process Meter
  • Instructions manual
  • TL71 Red/black test leads with AC72 test probes
  • Crocodile clips
  • 1 x 9V battery
Specification

DC Voltage

  • Ranges: 4V / 40V / 400V / 1000V
  • Resolutions: 0.001V / 0.01V / 0.1V / 1V
  • Accuracy: 0.1% + 1 / 0.1% + 1 / 0.1% + 1 / 0.1% + 1

DC Millivolts

  • Range: 400mV
  • Resolution: 0.1mV
  • Accuracy: 0.1% + 1

AC Voltage

  • Ranges: 400mV / 4V / 40V / 400V / 1000V
  • Resolutions: 0.1mV / 0.001V / 0.01V / 0.1V / 1V
  • Accuracy (50Hz to 60Hz): 0.7% + 4 / 0.7% + 2 / 0.7% + 2 / 0.7% + 2 / 0.7% + 2
  • Accuracy (45Hz to 200Hz): 1.5% + 4 / 1.5% + 4 / 1.5% + 4 / 1.5% + 4 / 1.5% + 4
  • Accuracy (200Hz to 500Hz): 7% + 4 / 7% + 4 / 7% + 4 / 7% + 4 / 7% + 4

AC Current

  • Range (45Hz to 2kHz): 0 to 1A*
  • Resolution: 0.001A
  • Accuracy: 1% + 2
  • Typical burden voltage: 1.5V/A

*440mA continuous, 1A for 30s max followed by 5 minute rest

DC Current

  • Ranges: 30mA / 1A*
  • Resolutions: 0.001mA / 0.001A
  • Accuracy: 0.05% + 2 / 0.2% + 2
  • Typical burden voltage: 14mV/mA / 1.5V/A

*440mA continuous, 1A for 30s max followed by 5 minute rest

Resistance

  • Range: 400Ω / 4kΩ / 40kΩ / 400kΩ / 4MΩ / 40MΩ
  • Resolution: 0.1Ω / 0.001kΩ / 0.01kΩ / 0.1kΩ / 0.001MΩ / 0.01MΩ
  • Measurement current: 220µA / 59µA / 5.9µA / 590nA / 220nA / 22nA
  • Accuracy: 0.2% + 2 / 0.2% + 1 / 0.2% + 1 / 0.2% + 1 / 0.35% + 3 / 2.5% + 3

Frequency Accuracy

  • Ranges: 199.99Hz / 1999.9Hz / 19.999kHz
  • Resolution: 0.01Hz / 0.1Hz / 0.001kHz
  • Accuracy: 0.005% + 1 / 0.005% + 1 / 0.005% + 1

Frequency Sensitivity

  • Input range: 1V / 4V / 40V / 400V / 1000V
  • Minimum sensitivity (RMS Sinewave) 5Hz to 5kHz: 0.1V / 1V / 3V / 30V / 300V

Diode Test and Continuity Test

  • Diode test indication: display voltage drop: 0.2mA nominal test current at 0.6V: 2.4V full scale, accuracy ±(2% + 1)
  • Continuity test indication: Continuous audible tone for test resistance <100Ω
  • Open circuit voltage: <3.9V
  • Short circuit current: 1.2mA typical
  • Overload protection: 1000V RMS

DC Current Output

Source mode

  • Span: 0mA or 4mA to 20mA, with overrange to 24mA
  • Accuracy: 0.05% of span
  • Compliance voltage: 12V with battery voltage >8.5V

Simulate Mode

  • Span: 0mA or 4mA to 20mA, with overrange to 24mA
  • Accuracy: 0.05% of span
  • Loop voltage: 24V nominal, 30V maximum, 15V minimum
  • Compliance voltage: 21V for 24V supply
  • Burden voltage: <3V

General

  • Max voltage: 1000V
  • Storage temperature: -40˚C to 60˚C
  • Operating temperature: -20˚C to 55˚C
  • Operating altitude: 2000m maximum
  • Temperature coefficient: 0.05 x specified accuracy per ˚C for temperatures <18˚C or >28˚C
  • Relative humidity: 95% up to 30˚C; 75% up to 40˚C; 45% up to 50˚C; 35% up to 55˚C
  • Vibration: Random Mil-prf-28800f, 10Hz to 500Hz
  • Shock: 1m drop test
  • Power requirements: 1 x 9V battery
  • Size: 32mm x 87mm x 187mm, 52mm x 98mm x 201mm (with holster)
  • Weight: 369g, 638g (with holster)
  • Safety: IEC 61010-1 1000V Cat III