Keller LEO1 Ei Intrinsically Safe Digital Manometer (700bar)

SKU: KELL303005.0009

Never Beaten On Price

 

We’re always working hard to give you the best possible price, but if you find it cheaper somewhere else, we’ll endeavour to match the price.

 

Call 0113 248 9966 or contact us online

Typically ships in 2-4 weeks
Recommended Add-ons
Need Some Help From Our Experts? chevron-right chevron-down

We're always happy to offer help, information & advice to find the right instrument for you, so please get in touch.

 

To speak to our sales team, call us on 0113 248 9966, use our Live Chat service or contact us online.

Members earn 385 reward points chevron-right chevron-down

Use reward points to get money off future purchases.

 

Please note: Customers must be signed into their account to be elegible. Find out more about reward points here.

FREE Delivery & 60 Day Returns chevron-right chevron-down

We offer FREE standard delivery on all orders over £249, with a range of next-day options also available on many items.

 

Find out more on our delivery information page.

Key Features
  • Pressure measurement from 0bar to 700bar with accuracy of ±0.1% FS
  • Classified as an intrinsically safe manometer for use in dangerous environments
  • PEAK mode takes 5000 readings per second and shows peak/trough measurements

Related Categories

Keller LEO1 Ei Intrinsically Safe Digital Manometer (700bar)

With a pressure measurement range of 0bar to 700bar, the Keller LEO1 Ei Intrinsically Safe Digital Manometer is capable of measuring the high pressures found in many industrial environments. It has an accuracy of ±0.1% FS and a resolution of 200mbar and takes measurements through a G ¼” male connector.

The Keller LEO1 Ei is classed as intrinsically safe for use in conditions where there is a risk of explosion such as mines and oil rigs. It has a classification of Ex II 1 G and EEx ia IIC T5 or T6 and a certification file of LCIE 01 ATEX 6001 X.

The Keller LEO1 Ei has two measurement modes. Firstly, ‘MANO’ mode takes two measurement per second and displays the minimum or maximum pressure alongside the current pressure reading. Secondly, ‘PEAK’ mode takes up to 5000 readings per second and shows trough and peak measurements with the current pressure reading.

A zero function on the Keller LEO1 Ei enables relative pressure measurement that is not affected by changes in barometric pressure. Readings are shown in multiple units on the clear digital display and the pressure gauge is operated using just two buttons.

The Keller LEO1 Ei has a battery life of 1000 hours in MANO mode and has an IP65 ingress protection rating so it is ideal for use on large sites.

Keller part number: 303005.0009

Key Features

  • Intrinsically safe
  • Suitable for ATEX environments
  • Classification: Ex II 1 G, EEx ia IIC T5 or T6
  • Measurement range: 0 to 700bar
  • Accuracy: ±0.1% FS
  • Resolution: 200mbar
  • Relative pressure measurement
  • PEAK mode takes 5000 readings per second
  • 1000-hour battery life
  ECO2 LEO1 LEO2 LEO3 LEO5 LEO Record LEX1
-1 to 3bar   ✔ ✔ ✔ ✔ ✔ ✔
-1 to 31bar ✔ ✔ ✔ ✔ ✔ ✔ ✔
0 to 100bar         ✔    
0 to 300bar ✔ ✔ ✔ ✔ ✔ ✔ ✔
0 to 400bar             ✔
0 to 700bar   ✔ ✔ ✔ ✔ ✔  
0 to 1000bar   ✔   ✔   ✔ ✔
Accuracy ±0.5% FS ±0.1% FS ±0.1% FS ±0.1% FS ±0.05% FS ±0.1% ±0.05% FS
Resolution
(at smallest range)
10mbar 1mbar 1mbar 1mbar 1mbar 1mbar 0.1mbar
Temperature range         0˚C to 50˚C    
Temperature accuracy         ±1˚C    
Intrinsically safe
version
✔ ✔ ✔   ✔ ✔  
Peak mode   ✔     ✔    
Pressure connector 7/16" UNF G 1/4" 7/16" UNF G 1/4" G 1/4" G 1/4" G 1/4"
Analogue output       ✔ ✔    
Internal memory         ✔ ✔ ✔
Data port       RS485 Binder USB RS485 Fischer RS485 Fischer
Compatible with
Keller-Bus protocol
      ✔ ✔   ✔
Show Trustpilot Reviews
Specification
No
  • Classification: Ex II G, EEx ia IIC T6 or T5
  • Certification file: LCIE 01 ATEX 6001 X
  • Range: 0 to 700bar
  • Resolution: 200mbar
  • Overpressure: 700bar
  • Accuracy (at room temperature): <0.1% FS
  • Total error band (0…50˚C): <0.2% FS
  • Storage temperature: -20 to 70˚C
  • Operating temperature: 0 to 50˚C
  • Compensated temperature range: 0 to 50˚C
  • Supply: 3V battery, CR2430
  • Battery life: 1000 hours in MANO mode, 150 hours in PEAK mode
  • Pressure connection: G1/4”
  • Ingress protection rating: IP65
  • Dimensions: 76mm x 118mm x 42mm
  • Weight: 210g
Label Type Size Download
pdf Keller LEO1 Datasheet pdf 472.2 KB Download