Keller LEO2 Ei Intrinsically Safe Digital Manometer (700bar)

SKU: KELL303010.0008
  • Intrinsically safe, ideal for use in combustion-prone environments
  • Pressure range of 0 to 700bar with ±0.1% FS accuracy
  • Min/max pressure measurement recording and zero function
More Info
In Stock - Available Next Day

Keller LEO2 Ei Intrinsically Safe Digital Manometer (700bar)

The Keller LEO2 Ei Intrinsically Safe Digital Manometer has a pressure measurement range of 0bar to 700bar with accuracy of ±0.1% FS and a resolution of 200mbar. Pressure is measured through the gauge’s 7/16” UNF connector and piezoresistive pressure transducer.

As an intrinsically safe pressure meter, the Keller LEO2 Ei can be used in hazardous conditions such as those found on oil rigs, down mines and in certain types of factory. It has a classification of Ex II 1G and EEx ia IIC T6 or T5 and the certification file is LCIE 01 ATEX 6001 X.

The Keller LEO2 Ei has a digital screen that clearly displays readings in a number of different units. Min/max recording is available and the pressure gauge also has a zero function that allows relative measurement, thus negating any effects of variable barometric pressure.

The Keller LEO2 Ei is supplied with a G1/4” adaptor and an O-ring swivel seal that allows the gauge to rotate through 360˚ and maintains a seal capable of withstanding up to 300bar. A battery life of 1000 hours and IP65 ingress protection rating make it ideal for tricky conditions.

Keller part number: 303010.0008

Key Features

  • Intrinsically safe for ATEX environments
  • Ex II 1 G; EEx ia IIC T6 or T5
  • Range: 0 to 700bar
  • Accuracy: ±0.1% FS
  • Resolution: 200mbar
  • Min/max pressure recording
  • O-ring seal and G ¼” adaptor
  • 7/16” UNF 20 male connector
  • 1000-hour battery life
  ECO2 LEO1 LEO2 LEO3 LEO5 LEO Record LEX1
-1 to 3bar  
-1 to 31bar
0 to 100bar            
0 to 300bar
0 to 400bar            
0 to 700bar    
0 to 1000bar      
Accuracy ±0.5% FS ±0.1% FS ±0.1% FS ±0.1% FS ±0.05% FS ±0.1% ±0.05% FS
Resolution
(at smallest range)
10mbar 1mbar 1mbar 1mbar 1mbar 1mbar 0.1mbar
Temperature range         0˚C to 50˚C    
Temperature accuracy         ±1˚C    
Intrinsically safe
version
   
Peak mode          
Pressure connector 7/16" UNF G 1/4" 7/16" UNF G 1/4" G 1/4" G 1/4" G 1/4"
Analogue output          
Internal memory        
Data port       RS485 Binder USB RS485 Fischer RS485 Fischer
Compatible with
Keller-Bus protocol
       
Specification
  • Classification: Ex II 1 G; EEx ia IIC T6 or T5
  • Certification file: LCIE 01 ATEX 6001 X
  • Range: 0bar to 700bar
  • Resolution: 200mbar
  • Overpressure: 700bar
  • Accuracy: <0.1% FS
  • Total error band: <0.2% FS
  • Storage temperature: -20˚C to 70˚C
  • Operating temperature: 0˚C to 50˚C
  • Compensated temperature range: 0˚C to 50˚C
  • Supply: 3V battery, type CR 2430
  • Battery life: 1000 hours’ continuous operation
  • Pressure connection: 7/16”-20 UNF (adaptor G ¼” included)
  • Protection rating: IP65
  • Dimensions: 59mm x 95mm x 32mm
  • Weight: 125g
Label Type Size Download