Keller LEX1 (1000bar) Digital Precision Manometer

SKU: KELL303030.0006

Never Beaten On Price

 

We’re always working hard to give you the best possible price, but if you find it cheaper somewhere else, we’ll endeavour to match the price.

 

Call 0113 248 9966 or contact us online

Typically ships in 5-7 working days
Need Some Help From Our Experts? chevron-right chevron-down

We're always happy to offer help, information & advice to find the right instrument for you, so please get in touch.

 

To speak to our sales team, call us on 0113 248 9966, use our Live Chat service or contact us online.

Members earn 760 reward points chevron-right chevron-down

Use reward points to get money off future purchases.

 

Please note: Customers must be signed into their account to be elegible. Find out more about reward points here.

FREE Delivery & 60 Day Returns chevron-right chevron-down

We offer FREE standard delivery on all orders over £249, with a range of next-day options also available on many items.

 

Find out more on our delivery information page.

Key Features
  • Measures absolute pressure up to 1000bar with excellent ±0.05% FS accuracy
  • Works in a Keller-Bus monitoring protocol with up to 128 instruments
  • Suitable for applications in laboratories and specialist facilities

Related Categories

Keller LEX1 Digital Precision Manometer (1000bar)

The Keller LEX1 Digital Precision Manometer is capable of measuring pressure from 0 to 1000bar with accuracy of ±0.05% FS and a resolution of 100mbar, which makes it suitable for use in calibration facilities, laboratories and specialist industrial applications. It takes measurements with a piezoresistive transducer in oil-filled casing via a G ¼” male pressure connector.

The Keller LEX1 has a clear digital screen that can show pressure readings in a number of different units. There is a min/max recording function and a zero function that enables relative measurement. A digital calibration feature with an end point and zero point helps maintain accuracy.

An RS485 Fischer data connection allows the Keller LEX1 to act as part of a wider monitoring system. An optional Keller K114A Cable Adaptor can be used to adapt the connection to a USB port. Free Keller CCS30 software is available to download from Keller’s website and can be used to coordinate a Keller-Bus monitoring protocol with up to 128 instruments.

With just two buttons, the Keller LEX1 is surprisingly easy to operate and it also has an excellent 2000-hour battery life. The IP65 ingress protection rating certifies that it is suitable for challenging environments.

Keller part number: 303030.0006

Key Features

  • Piezoresistive transducer in oil-filled housing
  • Absolute pressure range: 0 to 700bar
  • Accuracy: ±0.05%FS
  • Resolution: 100mbar
  • G ¼” male pressure connector
  • Suitable for labs and calibration centres
  • Different units of pressure
  • Digital calibration function
  • Works as part of a Keller-Bus protocol
  ECO2 LEO1 LEO2 LEO3 LEO5 LEO Record LEX1
-1 to 3bar ✔ ✔ ✔ ✔ ✔ ✔
-1 to 30bar ✔ ✔ ✔ ✔ ✔ ✔ ✔
0 to 100bar ✔
0 to 300bar ✔ ✔ ✔ ✔ ✔ ✔ ✔
0 to 400bar ✔
0 to 700bar ✔ ✔ ✔ ✔ ✔
0 to 1000bar ✔ ✔ ✔ ✔
Accuracy ±0.5% FS ±0.1% FS ±0.1% FS ±0.1% FS ±0.05% FS ±0.1% ±0.05% FS
Resolution(at smallest range) 10mbar 1mbar 1mbar 1mbar 1mbar 1mbar 0.1mbar
Temperature range 0˚C to 50˚C
Temperature accuracy ±1˚C
Intrinsically safeversion ✔ ✔ ✔ ✔ ✔
Peak mode ✔ ✔
Pressure connector 7/16" UNF G 1/4" 7/16" UNF G 1/4" G 1/4" G 1/4" G 1/4"
Analogue output ✔ ✔
Internal memory ✔ ✔ ✔
Data port RS485 Binder USB RS485 Fischer RS485 Fischer
Compatible withKeller-Bus protocol ✔ ✔ ✔
Show Trustpilot Reviews
Specification
No

Performance

  • Absolute pressure range: 0 to 1000bar
  • Overpressure: 1100bar
  • Resolution: 100mbar
  • Accuracy @ RT (20˚C to 25˚C): ≤±0.05% FS. Nonlinearity (BFSL), pressure hysteresis, non-repeatability, zero point, amplification.
  • Total error band (0˚C to 50˚C): ≤±0.05% FS. Max. deviation within the specified pressure and temperature range.
  • Long-term stability: ±0.1% FS. > 1 bar, per year under reference conditions, yearly recalibration recommended.
  • Long-term stability: ±0.1mbar. ≤ 1 bar, per year under reference conditions, yearly recalibration recommended.
  • Degree of dependency on location: <±1.5mbar. Calibrated in vertical installation position with pressure connection facing downwards.
  • Accuracy of temperature measurement: ±1˚C typ.
  • Pressure range reserve: ±10%

Electrical Data

  • Battery: 3 V, type CR2430. LEX1 Ei only permitted for use in haz- ardous atmospheres when used with the Renata CR2430
  • Battery life: Approx. 2000 hours when used continuously
  • External power supply: 8...28 VDC
  • Overvoltage protection and reverse polarity of external power supply: ± 32 VDC
  • RS485 voltage insulation: -7...+12 VDC
  • GND case insulation: > 10 MΩ @ 50 VDC
  • External interface: RS485 half-duplex
  • Interface measuring rate: 30 measurements per second
  • Electrical connection: Fischer D 103 A054-130
  • CE conformity as per 2014/30/EU (EMC): EN 61000-6-1 to -6-4; EN 61326-1 / EN 61326-2-3

Temperature Range

  • Compensated temperature range: 0...50 °C
  • Media temperature: 0...50 °C
  • Ambient temperature: 0...50 °C
  • Storage temperature: -20...70 °C

Display

  • Dimensions/appearance: See Dimensions & options
  • Number of digits on LC display: 2 rows with 5 digits each
  • Display mode: Pressure + min/max or pressure + temperature
  • Measuring rate: 2 measurements per second
  • Configurable units of pressure: [bar], [mbar], [hPa], [kPa], [MPa], [PSI], [mH2O], [cmH2O], [inH2O], [ftH2O], [mmHg], [inHg], [kp/cm2]
  • Additional units of pressure: 5 user-defined units can be configured

Mechanical data

  • Pressure connection: Stainless steel AISI 316L
  • Pressure transducer separating diaphragm: Stainless steel AISI 316L
  • Pressure transducer seal (internal): FKM (Viton® type A)
  • Pressure connection seal (external): FKM (Viton® type A)
  • Display housing: Faradex AS-1003
  • Oil filling sensor: Silicone oil
  • Pressure connection: G1/4, see Dimensions & options
  • Diameter x height x depth (approx.): 76 x 118 x 55 mm
  • Weight (approx.): 300 g
  • Protection: IP65
Label Type Size Download