  • Four channels for thermocouples including types K, J, E, T, N, R & S
  • Measuring range of -200˚C to 1700˚C with max resolution of 0.1˚C
  • Downloadable internal memory with min/max/avg reading function
More Info
Typically ships in 1-3 working days

Martindale DTL84 Multi-Input Data Logging Thermometer

The Martindale DTL84 Multi-Input Data Logging Thermometer has four channels compatible with a variety of thermocouples (type K, J, E, T, N, R, S). It can measure temperature from -200˚C to 1700˚C depending on the probe and accuracy can be as good as ±(0.1% + 0.7˚C) with a type J, K, E, T or N probe. Readings have a maximum resolution of 0.1˚C.

Four temperature readings can be shown simultaneously on the Martindale DTL84’s large digital screen in either Fahrenheit or Celsius and there is a data hold function that freezes a reading on screen. Additionally, a min/max/avg recording function allows the user to view extreme and average values as necessary, including when measuring over a longer period.

As a data logging instrument, the Martindale DTL84 is capable of storing results in its onboard memory as a spot reading or as part of longer-term monitoring. Data transfers can be performed via the USB port to a PC or by using the supplied Bluetooth adaptor to form a wireless connection with an iOS or Android smart device (app required). A data transfer software package is provided to facilitate manual downloads.

As standard, the Martindale DTL84 comes supplied with two type K thermocouples that have a temperature range of -50˚C to 200˚C and a probe accuracy of ±2.2˚C. Further accessories include the Bluetooth adaptor, a micro USB cable, 4 x 1.5V AA batteries, a software CD, a storage case and an instruction manual.

Martindale part number: MARDTL84

Key Features

  • Four-channel digital thermometer
  • Thermocouple compatibility: Type K, J, E, T, N, R, S
  • Measuring range: -200˚C to 1700˚C
  • Max. accuracy: ±(0.1% + 0.7˚C)
  • Max. resolution: 0.1˚C / 0.1˚F
  • Min/max/avg recording
  • Temperature monitoring for extended periods
  • Data transfer via USB or Bluetooth
  • Supplied with two type K thermocouples
  • Two-year warranty

What's In The Box?

  • Martindale DTL84 Multi-Input Data Logging Thermometer
  • 2 x type K thermocouples
  • Bluetooth adaptor
  • 4 x 1.5V AAA batteries
  • DTL84 software CD
  • Micro USB cable
  • Instructions for Martindale DTL84
Specification

All specified accuracies are at 23°C ± 5°C < 90% R.H. (non-condensing) for 1 year.

Temperature coefficient: 0.01% of reading + 0.05°C (0.0028°F) per °C (0°C to 18°C, 28°C to 50°C or 32°F to 64°F, 82°F to 122°F)

Ranges

  • K: -200°C-1372°C (-328°F-2501°F)
  • J: -200°C-1000°C (-328°F-1832°F)
  • E: -200°C-750°C (-328°F-1382°F)
  • T: -200°C-400°C (-328°F-752°F)
  • N: -200°C-1300°C (-328°F-2372°F)
  • R, S: 0°C-1767°C (32°F-3212°F)

Resolutions

  • K-, J-, E-, T-, N-type: 0.1°C < 600°C; 0.1°F < 1000°F; 1°C ≥ 600°C; 1°F ≥ 1000°F
  • R-, S-type: 0.2°C < 600°C; 0.5°F < 1000°F; 1°C ≥ 600°C; 1°F ≥ 1000°F

Accuracy:

  • J, K, E, T, N-type: -200°C to -100°C ± (0.5% of reading +0.7°C); 100°C ± (0.1% of reading +0.7°C); -328°F to -148°F ± (0.5% of reading +1.3°F); ≥ -148°F ± (0.1% of reading +1.3°F)
  • R, S-type: 0°C to 1767°C ± (0.2% of reading +1.4°C) 32°F to 3212°F ± (0.2% of reading +2.5°F)

General purpose 100cm Type K thermocouple bead probe

  • Temperature measurement range: -50°C to 200°C (-58°F to 392°F)
  • Probe accuracy: ± 2.2°C (± 3.9°F)

GENERAL

  • Display: Liquid crystal display
  • Over range: OL is displayed
  • Measurement rate: twice per second
  • Power: 4 x 1.5V AAA alkaline batteries (IEC LR03, NEDA 24A)
  • Continuous operating time: Bluetooth OFF – 120 hours approx; Bluetooth ON – 30 hours approx
  • Auto power off: Setable to 10mins, 30mins, 1hr, 2hrs, 4hrs, 8hrs or disabled
  • Dimensions: 187mm (L) x 75mm (W) x 29mm (H)
  • Weight: 290g
  • Includes: storage case, 2 x K-type thermocouples, bluetooth adaptor, 4 x 1.5V AAA alkaline batteries, DTL84 software CD, micro USB cable and instruction manual.

ENVIRONMENTAL

  • Temperature & Humidity: Operating: 0°C to 50°C, 10% to 90% R.H. non-condensing
  • Storage: -20°C to 60%, 10 to 75% R.H.

EMC

  • Conforms to BS EN 61326-1